Noshok
820/821 Series Digital Temperature Indicators

Key Features

  • Large 4-digit LED display
  • Field re-programmable with optional PC interface module and software
  • Software includes a security feature to prevent accidental re-programming
  • 4 mA to 20 mA programmable linearized signal output
  • Utilizes self-calibration feature for accurate and stable performance
  • Optional fully programmable switch output, relay or transistor
  • Utilizes PT100 ? RTD Class A element for temperature sensing
  • M12 x 1 (5-pin) plug or 36” integral cable electrical connection
  • IP 65 / NEMA 4 rated environmental protection
  • All 316 stainless steel construction
  • Easy installation with various mounting configurations
Click to Enlarge

Application:

  • Replacement for bimetal, liquid bulb and glass thermometers
  • Pharmaceutical
  • Food preparation
  • Utilities and municipal refineries
  • Chemical and petrochemical plants
  • Paper mills
  • Hydraulics

SPECIFICATIONS
Temperature ranges Standard ranges from -325 °F to 1,100 °F (-200 °C to 600 °C) Customer rescalable with optional PC interface and software
Temperature sensor RTD (PT100 DIN EN 60751, Class A)
Housing material 316 stainless steel
Probe material 316 stainless steel standard
Maximum pressure 500 psig (on probe)
Power supply 9-36 Vdc, polarity protected
Supply effect 0.005%/V
Power consumption 15 mA @ 24 Vdc + output current – 950 mW max.
20 mA @ 24 Vdc for PNP output – 500 mW max.
20 mA @ 24 Vdc + sourcing current for NPN output
50 mA @ 24 Vdc for relay output – 1200 mW max.
Current output 4 mA to 20 mA (3-wire configuration) linear to temperature
Max load on current output (Vsupply-9V ) / 20 mA, ?
Optional switching output Transistor NPN (max 100 mA source ) or
Transistor PNP (max 100 mA sink) or
Relay SPDT 0.5A @ 240 Vac
Optional switching logic N.C. or N.O. software selectable
Optional switching ranges Customer programmable between -325 °F to 1,100 °F (-200 °C to 600 °C)
Isolation 500 Vdc input /output (between probe and output signal)
Electrical connection M12 x 1 (5-pin) or integral cable
Hysteresis 1% of range standard; customer programmable optional
Accuracy ±0.25% full scale (0.45 °F + 0.40% of reading) max. with default calibration
±0.125% full scale (0.18 °F + 0.20% of reading) max. with one-point factory or customer calibration
Open circuit detection Upscale (22 mA) or downscale (2.5 mA) current output.
Error message on LED display
Warm-up 30 seconds
Response time 0.5 sec to 30 sec (software selectable)
Display 4-digit LED, decimal point selectable by software
Display resolution ±0.02% F.S. ±1 digit
RFI effect 1% or less typical
Temperature ranges Ambient -40 °F to 176 °F (-40 °C to 80 °C)
Effect <0.01% FS/ °C
Storage -58 °F to 185 °F (-50 °C to 85 °C)
Environmental protection NEMA 4/ IP 65